A402060121C ·
metrohm · Cat: A402060121C
Reach 3+ verified distributors
Manufacturer product page
metrohm.com/en_in/products/a/4020/a402060121c.htmlDescription
The 2060 Monitor for AeRosols and Gases in ambient Air MARGA uses ion chromatography to monitor air quality by measuring the concentration of water-soluble gases and aerosols in air containing different sizes of particulate matter. Air is drawn in to the instrument by a vacuum pump at a constant flow rate. Before entering the instrument, the maximum size of particulate matter PM can be limited by utilizing a size-selective inlet e.g. PM 1.0, PM 2.5 or PM 10 . The water-soluble gases are absorbed in a Wet Rotating Denuder WRD device and then the water-soluble ions in the aerosols are extracted in a Steam-Jet Aerosol Collector SJAC .A dual-channel ion chromatograph IC quantifies anions and cations in aerosols and water-soluble gases simultaneously with high accuracies to reach detection limits as low as 0.001 mug/m3.The analyzers performance can be remotely observed by using a direct internet connection to make adjustments when necessary and to access stored results. The 2060 MARGA R version is ideal for research campaigns: the IC can be used as a standalone lab instrument for other projects when not in use for 24/7 air quality monitoring.; Data collection with a time resolution of 1h to study e.g. diurnal variation or trajectory models for pollution events.; Remote control or touch screen to easily access results, trend graphs, and more.; Gases measured: Hydrochloric Acid HCl , Nitric acid HNO3 , Nitrous acid HONO , Sulfur dioxide SO2 , Ammonia NH3 ; Aerosol ions measured: Chloride Cl- , Nitrate NO3- , Sulfate SO42- , Ammonium NH4+ , Potassium K+ , Calcium Ca2+ , Magnesium Mg2+ ;